Cleaning And Maintenance; Manual Cleaning - BD EnCor EC180 Instrucciones De Uso

Sistema de biopsia mamaria, abrazadera del controlador 180
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 25
1.
Remove the standard Driver Bracket (B) from the EnCor
Biopsy Driver (G), if attached.
2.
Rotate the Dial (F) on the bottom of the EnCor
Driver (G) to expose the mounting holes. Attach the EnCor
Breast Biopsy Driver (G) to the EnCor
Bracket 180 (H) by aligning the Driver Mounting Pins (C) to
the EnCor
Mounting Screw (D). The EnCor
attaches to the Driver Bracket (B) in only one orientation.
3.
Verify that the EnCor
plate is contacting the shoulders of the Driver Mounting Pins (C)
and Driver Mounting Screw (D). Do not overtighten.
4.
Reassemble the EnCor
(B) to the EnCor
Driver Rail (E) until it is fully engaged and locked.
5.
Verify proper mounting and function by arming and triggering the
EnCor
Fire Forward Accessory (A).
Note: A disposable EnCor
6.
Refer to the Instructions for Use of the EnCor
Accessory for use of the EnCor
Note: To rotate the sampling chamber 180º from its current
orientation, hold down the "VAC" button while pressing the "SAMPLE"
button momentarily.

Cleaning and Maintenance

Note: The EnCor
intended to be sterilized. For Cleaning Instructions for the EnCor
Fire Forward Accessory, refer to the EnCor
Instructions for Use.

Manual Cleaning

After each use, immediately clean the EnCor
Bracket 180 (H) with the steps outlined below.
1.
Remove the EnCor
Fire Forward Accessory (A). Reference Figure 1.
2.
Thoroughly wipe all internal and external surfaces of the Driver
Bracket 180 (H) with a CaviWipes1 disinfectant towelette or
equivalent cleaning wipe. Do not remove the Driver Mounting
Screw (D) from the Driver Bracket 180 (H) (see Figure 1).
3.
For the steps mentioned above, ensure all surfaces of the
EnCor
Driver Bracket 180 (H) are wiped continuously and
remain wet with a contact time of 3 minutes minimum.
Breast Biopsy Driver (G) and tightening the Driver
Breast Biopsy Driver (G) internal mounting
Fire Forward Accessory (A) by sliding onto the
Breast Biopsy Probe is not required for
Breast Biopsy Driver Bracket 180 is not
Driver Bracket 180 (H) from the EnCor
Breast Biopsy Driver (G)
Breast Biopsy Driver (G)/ Driver Bracket
Fire Forward Accessory (A).
4
Breast Biopsy
Breast Biopsy Driver
Fire Forward
Fire Forward Accessory
Breast Biopsy Driver
Breast