Medical Network Adapters & Transcoders
2.4
Power connection (MNT-120)
Power connection – Equipment with external 12 VDC power supply
Power requirements: The equipment may be powered using the delivered medical approved 12 VDC (
The medical approved DC (
The power supply is specified as a part of the ME equipment or combination is specified as a ME system.
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
The equipment should be installed near an easily accessible outlet.
The equipment is intended for continuous operation.
2.5
Transient over-voltage (MNA-110, MNA-120, MNA-240 and MNT-120)
If the device is not used for a long time, disconnect it from the AC inlet to avoid damage by transient over-voltage.
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
2.6
Power cords (MNA-110, MNA-120, MNA-240 and MNT-120)
Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent, 18 AWG min., rated 250 V min., provided with a hos-
pital-grade type plug 5-15P configuration for 120V application, or 6-15P for 240V application.
Do not overload wall outlets and extension cords as this may result in fire or electric shock.
Mains lead protection (U.S.: Power cord): Power cords should be routed so that they are not likely to be walked upon or pinched
by items placed upon or against them, paying particular attention to cords at plugs and receptacles.
The power supply cord should be replaced by the designated operator only at all time.
Use a power cord that matches the voltage of the power outlet, which has been approved and complies with the safety standard
of your particular country.
2.7

Other connections

Connections with other peripherals (MNA-110, MNA-120, MNA-240 and MNT-120)
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1 3rd ed. or Table BBB.201
of IEC 60601-1-1 for the medical electrical systems.
SFP+ connections (MNA-110, MNA-120 and MNA-240)
The equipment must be connected to the internet with SFP+ modules (SFP-RJ45 connectors) at the switch port.
2.8

Environmental safety

Degree of safety (flammable anesthetic mixture):
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
Liquids and moisture
Never expose the device to liquids or moisture.
Never use the device near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry tub or in a wet basement.