Table 4 Multi wavelength setup options (Continued)
Options
Description
Single Reading: A single reading is displayed after Read is
pressed.
Carousel 1 inch square: Optional measurement of carousel
Reading Mode
insert with up to 5 square cells.
Carousel 1 cm square: Optional measurement of carousel
insert with up to 7 square cells.
Recall
Recall saved measurement data, wavelength scans or time
measurement
courses, see
Save, recall, send and delete data, page
data
Basic data of the instrument, refer to
Instrument Setup
(absorbance, concentration and transmittance readings),
mode
page
161.
English 164
Wavelength Scan mode – recording of absorbance
and transmission spectrums
In the Wavelength Scan mode, the absorbance of the light in a solution
over a defined wavelength spectrum is measured.
The reading results can be displayed as a curve, as percentage
transmittance (%T) or as Absorbance (Abs). The collected data can be
printed as a table or a curve.
The data is available for formatting changes. These include automatic
scaling and zoom functions. Maximum and minimum values are
determined and shown as a table.
The cursor can be moved to any point on the curve for the purpose of
reading off the absorbance or transmittance value and the wavelength.
The data associated with each data point can also be shown as a table.