Now you have to find the cause of the light alarm on the low nursing bed sentida 7-i.
Example: A brake was released. Lock the brake once again.
▪
Then, turn off the light alarm in the alarm menu. To do so, press the crossed out speaker
symbol.
The red highlighted flashing speaker symbol is now turning green. This indicates the
proper condition of the monitored function.
If the light alarm cannot be transmitted to the call system despite a properly wor-
king CareBoard
switched „OFF". Switch „ON" the intermediate switch. If still no light alarm can be
transmitted, a technical defect on the intermediate switch, the NurseCall box or
the connection cable might be the case. In this event, please contact the wi-bo
customer service.
If a humidity sensor mat was not connected, the pictogram is crossed out in red.
Sensitivity
Change code
„Safe" mode: A call is triggered if any one of four raised side guards is lowered.
„Free" mode: A call is triggered if the fourth side guard is raised when three side guards were
already up.
88
, the intermediate switch on the adapter cable (if used) might be
®
Humidity assistant submenu
In this submenu, which you can open by pres-
sing the gear symbol next to the humidity as-
sistant, you can adjust the sensitivity of the
humidity sensors from 1 to 8 (8 = maximum
sensitivity).
Side guard assistant submenu
The submenu for the side guard assistant can
also be opened by pressing the corresponding
gear symbol. Here, you can select between
„Safe" and „Free" mode. To switch modes, you
have to enter a password in the field below.