Window-Based Protocol
The following table, valid for the RS 232, RS 422 and RS 485
interfaces, describes each single window used in the protocol.
WIN
000
001
100
101
102
103
104
105
106
107
108
109
120
121
200
201
202
203
205
206
207
208
Turbo-V 6000 Controller User Manual / 87-900-855-01
R
W
T
Description
X
X
L
START/STOP
X
X
L
LOW SPEED ON/OFF
X
X
L
SOFT START YES/NO
X
X
L
DEAT TIME YES/NO
X
X
L
WATER COOLING YES/NO
X
X
N
SPEED THRESHOLD
X
X
N
RUN UP TIME (0÷359.999 sec)
X
X
N
LOW SPEED ADJUST
X
X
N
HIGH SPEED ADJUST
X
X
N
MODE (0, 1, 2) [FRONT, REMOTE, SERIAL]
X
X
N
BAUD_RATE (0-4) [600, 1200, 2400, 4800, 9600]
X
L
PUMP LIFE RESET (TYPE "ON" TO RESET)
X
X
L
EXTERNAL INPUT READ ENABLE [YES/NO]
X
X
L
HEATER BAND ENABLE [YES/NO]
X
N
CURRENT [ A]
X
N
VOLTAGE [ V]
X
N
POWER [ W]
X
N
EXCITEMENT FREQUENCY [ Hz]
X
N
PUMP STATE (0÷6) [STOP, WAITING INTERLOCK,
STARTING, NORMAL, HIGH LOAD,
FAILURE,APPROACHING]
X
N
ERROR CODE (0÷8) [NO ERROR, OVERVOLTAGE,
SHORT CIRCUIT, CHECK CONNECTION, TOO HIGH
LOAD, OVERRIDE, PUMP OVERTEMP, CONTROLLER
OVERTEMP, OIL LEVEL AT MIN]
X
L
STATE R1 ON/OFF
X
L
STATE R2 ON/OFF
Technical Information
Window-Based Protocol
(F)
17