Voltage testers of P type are used to test voltage,
continuity of connections, diode tests, checking
phase rotation, and - in some models - resistance
measurements and identification of phases.
In order to provide conditions for correct operation
and the correctness of the obtained results, the fol-
lowing recommendations must be observed.
Before you proceed to operate the device, ac-
quaint yourself thoroughly with this manual and
observe the safety regulations and specifications
defined by the producer
Any application that differs from those specified
in the present manual may result in a damage to
the device and constitute a source of danger for
the user.
The device should be operated only by suitably
qualified persons having the necessary permis-
sions to carry out measurements on electrical
systems. Unauthorized use of the tester may re-
sult in its damage and may be a source of seri-
ous hazard to the user.
Comply with safety requirements and regulations
in force in your country when working with the de-
vice. This also applies to the use of personal pro-
tective equipment to protect against electric
shock.
Before using the device, please check its accura-
cy on the known voltage source.
Do not perform measurements in explosive at-
mosphere (e.g. in the presence of flammable
gases, vapours, dusts, etc.). Using the meter in
such conditions may result in sparking and cause
an explosion.
It is unacceptable to operate the device when:
 it is damaged (including cracks and missing
parts of the housing) and completely or par-
tially out of order,
 its cable insulation is damaged,
 it was stored for an excessive period of time
in disadvantageous conditions (e.g. exces-
sive humidity)After moving the device from
a cool to a warm place with a high level of
relative humidity, do not start measure-
ments until the device is warmed up to the
ambient temperature (approximately 30
minutes)
Do not use the tester when the battery compart-
ment is removed or open. Do not power the me-
ter from other sources than those mentioned in
this manual.
When the low battery indicator lights up, test re-
sults may be incorrect.
Do not perform voltage measurements for longer
than 30 seconds. After the measurement period
of 30 s, the next measurement may be made no
earlier than after 240 s.