3.8
Phase sequence test
The function is active only with efficient
batteries.
The measurement may be performed us-
ing gloves - the third pole is capacitively
coupled to test probe L2 held by the user.
Indicator LEDs and symbols L and R on
the display operate only during the meas-
urement of AC voltage, but the phase se-
quence may be determined only in three-
phase systems.
Connect the probe tip:
L1 to suspected phase L1,
L2 to suspected phase L2.
Firmly grip both probes L1 and
L2 (to achieve capacitive cou-
pling to the earth).
The value of phase-to-phase
voltage is shown by:
LED scale,
display
Signalling the phase sequence
by activating:
indicator lights,
icons on the display
Description of indicator lights
After switching the probes,
reversed symbol will be acti-
vated.
P-4 ● P-5 ● P-6 – USER MANUAL
clockwise sequence
[Right]
(phase L2 precedes L1)
counter-clockwise se-
quence [Left]
(phase L1 precedes L2)
.
31
.