7. Patient evaluation

Read Working Instructions
Patients who are electrically-susceptible (patients with exposed, non-waterproof electric leads,
monitors, etc.), patients carrying infectious disease, or patients with certain other medical, mental or
physical conditions should not receive treatment with the whirlpool.
The trained operator must evaluate and verify that each patient is suitable for hydrotherapy treatment
before permitting the patient to begin hydrotherapy. If in doubt, consult a medical professional before
providing hydrotherapy treatment.
Read Working Instructions
To reduce the risk of exposure to bloodborne diseases such as VIH-1 and hepatitis when using the
whirlpool, read and follow disinfecting and cleaning instructions in this manual thoroughly.

8. Inspecting the whirlpool bath

This product has been carefully packaged at the factory to minimize the possibility of damage during
shipping.
Inspect the packaging for external signs of damage.
Inspect the contents for damage.
If there is visible damage to the instrument upon receipt, inform the shipping company and Ortotecsa
immediately.
Read Working Instructions
Do not attempt to operate this equipment if there is evidence of shipping damage or you suspect the
unit is damaged. Damaged equipment may present additional hazards to you. Contact Ortotecsa
technical support for advice before attempting to plug in and operate damaged equipment.
Have your facility's equipment maintenance personnel inspect the whirlpool regularly. Follow the
checklist and operate the whirlpool through all its functions as described in this manual.
Inspection Check List
1.
Are all components present?
2.
Is the whirlpool free of excessive wear?
3.
Does the turbine mount securely in place?
4.
Is a properly-grounded and voltage-matched hospital grade receptacle available for the turbine?
5.
Is the outlet or turbine cord equipped at the desired Height?
6.
Does the drain valve open and close properly?
7.
Is the thermometer present and is legible?
8.
Do the installed accessories operate without interfering with whirlpool bath use or turbine operation?