Instructions for Use

Use

Use

This paragraph details the fundamental operating procedures.
Make all electrical an pneumatic connections before the use of the
system.
WARNING!
Never use the turbopump when the inlet flange is not connected to the
vacuum chamber or is not blanked.
Do not touch the turbopump or any of its accessories during the heating
process. The high temperatures may cause burns.
Avoid impacts, oscillations or harsh movements of the pump when in operation.
CAUTION!
The bearings may become damaged.
Use air or inert gas free from dust or particles for venting the pump. The pressure
at the vent port must be less than 2 bar (above atmospheric pressure).
For pumping aggressive gases, these pumps are fitted with a special port to
allow a steady flow of inert gas (like N2, Ar) for pump bearing protection (see the
appendix "Technical Information").
When employing the pump for pumping toxic, flammable, or radioactive
WARNING!
gases, please follow the required procedures for each gas disposal.
Do not use the pump in presence of explosive gases.
168/224
All manuals and user guides at all-guides.com
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)