For Your Safety; General Safety Instructions - Dräger X-act 5000 Basic Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

For your safety

1
For your safety
1.1

General safety instructions

Strictly follow the instructions for use. The user must fully
understand and strictly observe these instructions. Use the
Dräger X-act 5000 Basic only for the purposes specified in
Chap. 2.3 on page 8.
In addition, follow the instructions for use of the Dräger-Tube
hereafter referred to as tube, as well as of the sampling tubes and
systems.
Do not dispose of the instructions for use. Ensure that they are
retained and appropriately used by the product user.
Only fully trained and competent users are permitted to use this
product.
Comply with all local and national rules and regulations associated
with this product (e.g. IEC 60079-14).
Only trained and competent personnel are permitted to inspect,
repair, and service the product as detailed in the Maintenance
section of this document. Dräger recommend a Dräger service
contract for all maintenance activities and that all repairs are
carried out by Dräger.
Only use genuine Dräger spare parts and accessories. Otherwise,
the proper functioning of the product may be impaired and there is
the risk of explosion. Furthermore, the approval will be invalid.
Do not substitute components of this product. Substitution of
components may impair intrinsic safety.
Do not use a faulty or incomplete product, and do not modify the
product.
Notify Dräger in the event of any component fault or failure.
1)
For US, AU, CA, GB Dräger-Tube
®
CH Dräger Röhrchen
are registered trademarks of Dräger.
4
®1)
TM
®
, for DK DRÄGERRØR
and for DE, AT,
Only use equipment or components which have been tested and
approved according to the national regulations on electrical
equipment in rooms subject to explosion hazards and according to
the national mining regulations under the conditions specified in
the approval.
Prior to safety-relevant measurements, perform a leak test and
check the calibration.
The operation modes of tubes and tube pumps are harmonized.
,
Using tube pumps and tubes from different manufacturers can
impair the proper function of the tube pumps and/or tubes. This
may result in considerable measurement errors.
Condition of safe use: The measured capacitance of the exposed
metal screws is 5 pF.
Dräger X-act 5000 Basic
Tabla de contenido
loading

Tabla de contenido