TECHNICAL DATA
Dial Position
Current AC
Current DC
Voltage AC
Voltage DC
Low Input Impedance
Hertz
Resistance
Continuity
Capacitance
Temperature
These instruments are True-RMS sensing. All voltage and current readings are True-RMS values.
Accuracy is specifi ed for 1 year after calibration, at operating temperatures of 18°C to 28°C, with relative humidity
at 0 % to 85%.
Input impedance Voltage AC
Input impedance Voltage DC
Overload protection
Voltage DC, voltage AC,
Lo-Z Voltage, Hertz
Current DC & Current AC:
Resistance, Continuity, Capacitance and
Temperature:
Maximum measurement time:
Minimum frquency measurement
Minimum AC Current measurement
For AC Voltage, AC Current and Lo-z:
Additional Accuracy by Crest Factor (C.F.):
Max. Crest Factor:
Measurement accuracy of square wave and
truncated waveforms at 1kHz is unspecifi ed.
Maximum voltage between any terminal and earth ground
Temperature Operating
Temperature Storage
Temperature Coeffi cient
Operating Altitude
Battery
Battery run time
Weight with battery
Safety Compliance
Temperature Range, Resolution and Accuracy are for the instrument. Temperature Probe may have different
specifi cations.
30
ENGLISH
414 012 - Multimeter 2217-40.indd 30
Range
6 A
10 A
6 A
10 A
600 mV / 6 V
60 V / 600 V
600 mV / 6 V /
60 V / 600 V
600 V
99.99 Hz / 999.9 Hz
9.999 kHz / 50.00 kHz
600 Ω / 6 kΩ / 60 kΩ / 600 kΩ
6 MΩ
40 MΩ
Cont Buzzer
0-600.0 Ω
100 μF
1000 μF
-40°C - 400°C
Resolution
0.001 A / 0.01 A
0.001 A / 0.01 A
0.1 mV / 0.001 V
0.01 V / 0.1 V
0.1 mV / 0.001 V
0.01 V / 0.1 V
0.1 V
0.01 Hz / 0.1 Hz /
0.001 kHz / 0.01 kHz
0.1Ω / 0.001kΩ / 0.01kΩ/ 0.1kΩ
0.001 MΩ
0.01 MΩ
0.1 μF
1 μF
0.1°C
1 minute at 10A, rest time 20 minutes minimum
Add 1.0% for C.F. 1.4 ~ 2.0, Add 2.5% for C.F. 2.0 ~ 2.5, Add 4.0% for C.F. 2.5 ~ 3.0
1.6 for 6600 ~ 5000 digits,2.0 for 5000 ~ 3000 digits, 3.0 for 3000 ~ 0 digits
0. 1 x specifi ed accuracy /°C (<18°C or >28°C)
CAN/CSA C22.2 NO.61010.1-2004; UL61010-1-2004;
Accuracy
±(1.5% +3dgt) (45-500Hz)
±(1.0% +3dgt)
±(1.0% +3dgt) (45-500Hz)
±(2.0% + 3dgt) (500-1000Hz)
±(0.5% +2dgt)
±(2.0% + 3dgt) DC,AC: 45-500Hz
±(0.1% + 2dgt)
Sensitivity: 10 Vp-p
±(1.0% +5dgt)
±(1.0% +5dgt)
±(2.0% +5dgt)
Buzzer sounds at 30Ω or less
±(1.9% +2dgt)
±(1.0% +10dgt)
10 MΩ / < 100 pF; Lo-z ≈ 4 kΩ
10 MΩ; Lo-z ≈ 4 kΩ
AC/DC 720V for 10 second
AC/DC 720V for 10 second
AC/DC 20A for 10 sec.
AC/DC 600V for 10 second
-10°C - 50°C
-40°C - 60°C
about 30 h with all functions
IEC/EN61010-1 2nd Edition;
Category III 600V;
Pollution Degree 2;
EMC EN61326-1
10.06.2010 14:14:57
2Hz
0.040A
600 V
2000 m
2 AA