Troubleshooting - Omron MicroAIR U100 Manual De Instrucciones

Ocultar thumbs Ver también para MicroAIR U100:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 130

Troubleshooting

In case of any of the below problems occur during use, first check that no other electrical device is
within 30cm.
If the problem persists, please refer to the table below.
Trouble
The nebulization rate is
extremely low.
The green power
indicator does not turn
on and the unit does not
nebulize.
Possible cause
Batteries are low (the orange
battery low indicator flashes).
The mesh is stained or clogged.
The batteries are depleted (the
orange battery low indicator turns
on).
The capacity of rechargeable
batteries is low.
The batteries are inserted in the
wrong direction.
The AC adapter (Optional) is not
connected correctly to a power
outlet.
The contact on the AC adapter
(Optional) is contaminated.
Corrective action
Replace the batteries with new
ones. Refer to page 12.
1. Clean the mesh by boiling. Refer
to page 22.
2. Replace the mesh cap with a new
one. Refer to page 14.
Replace the batteries with new ones
or use the AC adapter (Optional).
Refer to page 12 or 13.
Charge the batteries with a
commercially available recharger.
Re-insert the batteries in the correct
direction. Refer to page 12.
Check that the AC adapter is
connected to a power outlet and
main unit. Unplug then reinsert the
plug if necessary.
Wipe away any contamination with a
cloth that has been moistened with
water or sanitizing alcohol and then
wrung out lightly.
EN
25
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Ne-u100-e

Tabla de contenido