3. Connect the bipolar resectoscope with the Erbe bipolar cable
for resectoscopes to the RESECTOSCOPE socket (2).
You must use the adapter to access the optimized BIPOLAR CUT
++ and BIPOLAR SOFT COAG ++ modes. These modes are not
available if you directly connect the resectoscope to the MF sock-
et on the VIO.
You can only connect resectoscopes to the RESECTOSCOPE sock-
et. Other instruments fit in the socket, but do not work. If you
want to use the MF socket on the VIO with an instrument, remove
the adapter.
Ambient conditions for operation +10 °C to +40 °C.
5

Cleaning and disinfection

Do not clean in the washing machine!
Erbe recommends wiping or spray disinfection with an appropri-
ate disinfectant. Comply with the disinfectant manufacturer's
stipulations as to its suitability and concentration.
Do not sterilize!
6

Maintenance

Safety checks must be performed at least once a year. The safety
checks determine whether the safety and operational readiness
of the product conforms to a defined technical required status.
To carry out safety checks, read the service manual for the VIO D
devices.