19 Periodically check the battery for corrosion. Take out the
batteries from the oximeter if you do not expect to use it
within one month.
20 The
device
immediately after power-on to confirm all the display
segments and icons are shown and the speaker sounds
a few seconds tone. If you do not hear the POST pass
tone, it indicates the alarm system does not work well.
Please do not use the oximeter and contact qualified
service
representative.
21 Before using it, the user should carefully read the
applicable user manual of sensor, including warnings,
cautions and instructions.
22 Do not use damaged sensor or extension cables, do not
use sensor with exposed optical components.
23 Tissue damage may be caused by incorrect application
or prolonged measurement duration using the sensor
(more than 2 hours). Inspect the sensor periodically
according to sensor user manual.
24 Do not immerse or wet the sensor, as this may damage
the sensor.
25 There are no user-serviceable parts inside the oximeter,
the cover should only be removed by qualified service
personnel.
26 Do not spray, pour, spill liquid to the oximeter and its
accessories, connector, switch or opening in enclosure,
as this may damage the oximeter.
27 Before cleaning the oximeter or the sensor, make sure
that the equipment is switched off and disconnected
from the power line.
28 Reusable sensors must be moved to a new site at least
every 4 hours. Because individual skin condition affects
the ability of the skin to tolerate sensor placement. It
may be necessary to change the sensor site more
enters
personnel
or
POST
(Power-On-Self-Test)
your
local
8
manufacturer's