Product benefits

• An equipotential ring distributes the current and thereby the
heat to both halves of the neutral electrode nearly simulta-
neously.
• It is irrelevant in which direction you position the NESSY
Omega. The neutral electrode can be oriented universally.
• The contact gel overlaps all aluminum edges. This reduces
edge heating of the NESSY Omega.
• Skin-friendly, conformable fleece as carrier material.
• Pull-off label for documentation purposes.
• Pull-off tab to easily lift the neutral electrode from the pro-
tective film and skin.
• The product is suitable for latex allergy sufferers. It contains
no latex and no typical natural rubber additives, i.e. cannot
trigger any allergic reactions to latex.
4

Operating principle

The return electrode conducts a high-frequency current from the
patient to the electrosurgical unit across a large surface.
A split return electrode allows continuous monitoring of the re-
turn electrode with a suitable electrosurgical unit.