Use: Zero level of the red scale in dB corresponds to U
range. The reading is direct in dB for the 10 V
For the other ranges, it is possible to read in dB (near value) by adding respectively:
+10 dB on the 30 V
+20 dB on the 100 V
+30 dB on the 300 V
+40 dB on the 1000 V
DC AND AC CURRENTS
Always switch off the circuit. If the «Fus» light comes on, change the faulty
fuse(s) (Reminder: minimum voltage of 110 V).
Connect the leads to the multimeter and connect in series to the circuit with:
- the red lead in the «µmA» terminal, up to 1.5 mA
- the red lead in the «15 A» terminal, from 1.5 A to 15 A
Switch off the power supply to the circuit before changing range.
To get the current in µA, mA or A, multiply the value read on the appropriate scale
by the reading coefficient shown in the table.
A
Scale
Reading
coefficient
Voltage drop at
the terminals
(2)
Accuracy
(3)
Protection
Limitation 10 min. ON, 5 min. OFF up to +40°C max.
(1) Common to the 100 mV
(2) Without the leads. Resistance of the pair of leads supplied: approx. 70 mΩ.
(3) In % of the end of scale.
Do not use the A
A
(1)
Scale
Reading coefficient
Voltage drop at the
terminals
(2)
Accuracy
(3)
Protection
Limitation 10 min. ON, 5 min. OFF up to +40°C max.
(1) Bandwidth: 40 Hz to 5 kHz. See § 4.
(2) Without the leads. Resistance of the pair of leads supplied: approx. 70 mΩ.
(3) In % of the end of scale.
Presence of direct component give false measurement.
range
(1)
50 µA
150 µA
100
x0.5
x10
< 0.1 V
range.
input on unprotected current transformers.
1.5 mA
x0.1
range
range
range
1.5 mA
15 mA
x0.1
x1
< 0.25 V
2.5 %
Fuse 1.6 A HPC
15 mA
150 mA
15 red
x1
x10
< 0.8 V
5 %
Fuse 1.6 A HPC
14
= 0.775 V for the 10 V
0
range from 0 to +22 dB.
150 mA
1.5 A
15 black
x10
x0.1
< 0.3 V
1.5 A
x0.1
< 0.9 V
15 A
x1
< 0.8 V
5 %
Fuse 16 A HPC
15 A
x1
< 0.8 V
3 %
Fuse 16 A HPC