0207-050-702 Rev-L
Specifications
The SurgiLav Plus Irrigation System, REF 0207-050-000E is intended for use in the electromagnetic environment specified below.
The customer or the user of the SurgiLav Plus Irrigation System, REF 0207-050-000E should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
Not Applicable
IEC 61000-3-2
Voltage fluctuations/flicker
Not Applicable
emissions
IEC 61000-3-3
The SurgiLav Plus Irrigation System, REF 0207-050-000E is intended for use in the electromagnetic environment specified below.
The customer or the user of the SurgiLav Plus Irrigation System, REF 0207-050-000E should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
SurgiLav Plus Irrigation System, REF 0207-050-000E is used exceeds the applicable RF compliance level above, the SurgiLav Plus Irrigation System, REF 0207-050-000E should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the SurgiLav Plus Irrigation System, REF 0207-050-000E.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
www.stryker.com
Print Date: Aug 29, 2016 01:59:30 PM
0000004063, Rev. L Effective Date: -
Guidance and manufacturer's declaration - electromagnetic emissions
Group 1
The SurgiLav Plus Irrigation System, REF 0207-050-000E uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The SurgiLav Plus Irrigation System, REF 0207-050-000E is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply network that supplies buildings
Guidance and manufacturer's declaration - electromagnetic immunity
Compliance level
3 Vrms
3 Vrms
150 kHz to 80 MHz
3 V/m
3 V/m
80 MHz to 2.5 GHz
Electromagnetic environment - guidance
used for domestic purposes.
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to
any part of the SurgiLav Plus Irrigation System, REF 0207-050-000E, including
cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m)
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey
, should be less than the compliance level in each frequency range
a
Interference may occur in the vicinity of equipment marked with the following
(Non-ionizing electromagnetic radiation)
d=1.2√P
d=1.2√P
d=2.3√P
symbol:
EN
.
b
7