1.
S S o o a a k k
Prepare an enzymatic detergent according to the manufacturer instructions.
Fully immerse the device in the detergent. Use a syringe to inject 50 mL of the
detergent into any mated surfaces and crevices.
Soak the device in the detergent for 15 minutes.
2.
B B r r u u s s h h
Thoroughly brush the exterior of the device with a soft-bristle brush for 90 seconds,
focusing on any mated surfaces and crevices.
Use a syringe to inject 50 mL of the detergent into any mated surfaces and
crevices 5 times.
3.
R R i i n n s s e e
Remove the device from the prepared detergent. Rinse the device with reverse
osmosis/de-ionized (RO/DI) water at ambient temperature for 90 seconds or until
all visible detergent residue is removed.
Flush any mated surfaces and crevices 5 times. After all visible detergent residue
is removed, continue to rinse for 30 seconds.
Drain excess water from the device and dry it with a clean cloth or filtered
pressurized air.
Visually inspect the device for cleanliness, paying close attention to hard-to-reach
areas. If visible soil remains, repeat steps 1 through 3.
4.
S S o o a a k k
Prepare a non-enzymatic detergent according to the manufacturer instructions.
Fully immerse the device in the detergent. Use a syringe to inject 50 mL of the
detergent into any mated surfaces and crevices.
Soak the device in the detergent for 15 minutes.
5.
B B r r u u s s h h
Thoroughly brush the exterior of the device with a soft-bristle brush for 90 seconds,
focusing on any mated surfaces or crevices.
Use a syringe to inject 50 mL of the detergent into any mated surfaces and
crevices 5 times.
6.
R R i i n n s s e e
Remove the device from the prepared detergent. Rinse the device with reverse
osmosis/de-ionized (RO/DI) water at ambient temperature for 90 seconds or until
all visible detergent residue is removed.
Flush any mated surfaces and crevices 5 times. After all visible detergent residue
is removed, continue to rinse for 30 seconds.
Drain the excess water from the device and dry it with a clean cloth or filtered
pressurized air.
Visually inspect the device for cleanliness, paying close attention to hard-to-reach
areas. If visible soil remains, repeat steps 1 through 6.