Cleaning and Disinfection
NOTE: Refer to the disinfectant manufacturers' specifications for information about achieved levels of
disinfection.
Cleaning and Disinfection Schedule
The Adapter Cable exterior must be cleaned after every use, and then disinfected. Optionally, if you suspect
that the Adapter Cable's internal cavity has become contaminated by fluids, you can also clean and disinfect
the internal cavity:
Required, After Every Use
• Clean and disinfect the Adapter Cable exterior
Cleaning Agents and Disinfectants
Use only recommended cleaners and disinfectants listed below. Others may damage the Adapter Cable and
shorten product lifetime or cause safety hazards.
Tensides
•
Edisonite Schnellreiniger
*The Adapter Cable has been tested and qualified to withstand the following cleaning agents: Ethanol 70%
and Isopropanol 70%
Alcohol-Based
• Ethanol 70%
• Isopropanol 70%
®
• Incidin
Liquid
®
• Terralin
Liquid
*The Adapter Cable has been tested and qualified to withstand the following disinfectants: Ethanol 70%,
Isopropanol 70%, Cidex, and Cidex-Plus
Cleaning and Disinfecting Schedule
Recommended Cleaning Agents*
®
Recommended Disinfectants*
Aldehyde-Based
•
Metricide
•
Metricide-Plus
®
•
Cidex
•
Cidex-Plus
•
Kohrsolin
•
Omnicide
Optional, As Necessary
• Clean the Adapter Cable's internal cavity
• Disinfect the Adapter Cable's internal cavity
•
Ethanol 70-80%
•
Isopropanol 70-80%
•
Window Cleaner
®
®
®
®
®
3
Alcohols
Acid-Based
®
•
Mucapur