Screen description:
"%SpO
": SpO
symbol; "99": SpO
value;
2
2
2
"PR": Pulse rate icon; "65": Pulse rate value;
"
": Pulse beat symbol;
" ": Pulse intensity histogram.
"PI": Perfusion index icon; "1.4": Perfusion index value;
5. At this time, longtime press Display Key (about 2 seconds), display
screen will be shown as Figure 5. The PR is replaced by PI%. To return
to PR display, just longtime press Display key again. If there is no operation
in 20 seconds, it will return to PR display as well.
6. Change display direction.
Four directions display alternately. Press Display Key once at a time,
display screen will be flipped 90° each time in a cyclic manner. But there
is a little difference, the display facing left or right will not show
plethysmogram.
7. Alarm Indicator
When measuring, if SpO
value and pulse rate value exceeds the preset
2
alarm limits, the device will alarm automatically and the value on the screen
exceeding limit will blink; at this time press Display Key to shut down the
alarm. Exceeding pulse rate alarm limit: sound twice as an interval.
Preset alarm range:
SpO
alarm: Low limit: 90%
2
Pulse Rate alarm: High limit 120bpm - Low limit: 50bpm
7. Instructions for Charging the Cell
The rechargeable coin cell is used to supply power for the oximeter. When
it is run out, you need to use the provided charger to charge it, details is
as follows:
1. Insert the cell into charger; pay attention to its polarity (as shown in the
following figure).