Calibration - OHAUS EB serie Manual De Instrucciones

Ocultar thumbs Ver también para EB serie:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

6. CALIBRATION

For best results, calbrate the scale at regular ntervals. Temperature changes,
gravty varatons, alttude changes and abuse are few reasons why a scale may
need recalbraton.
When the scale s deally postoned for operaton, enter calbraton and proceed as
follows:
Long-press ZERO and COUNT at the same tme to ntate
calbraton. The requred calbraton weght flashes on the
dsplay (ex. 3000g).
At ths tme, a dfferent calbraton weght value can be selected by pressng .
Avalable selectons are 1/3, 2/3 and 100% of full capacty.
Once the desred value s dsplayed, place the correspondng weght on the pan.
The dsplay flashes untl the actual weght s regstered and the calbraton process
ends (the scale re-starts). At ths tme, remove the weght on the pan before the
dsplay count-down sequence ends.
NOTES:
• Calbraton weght s dsplayed n grams only.
• To abort the calbraton process, turn the scale off.
• If needed, the factory default calbraton value can be recovered. As the
scale s turned on, press TARE and COUNT at the same tme durng the power-on
sequence. The power-on sequence wll re-cycle and the scale wll rentate wth the
default calbraton.
EB SERIES EN-11
Tabla de contenido
loading

Este manual también es adecuado para:

Eb3Eb6Eb15Eb30

Tabla de contenido