Service And Maintenance; Troubleshooting - FINO DOSIL 14290 Información Sobre El Uso

Mezcladora y dosificadora
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 17
9.2.2 Mixing tip
Fix mixing tip at the mixing head (no. 4) by turning it 90° to the right.
The turning of the mixing tip can be eased by using the enclosed star-turn handle. The FINODOSIL Dosing
and Mixing Unit is now ready to operate.
9.2.3 Duplicating
Place your duplicating flask under the mixing tip and press the activation button (no. 2). As long as the
silicone in the mixing tip is not cured (approx. 4 minutes), an arbitrary number of flasks can be filled one after
another. Leave the tip on the mixing head after the filling process. It will serve as a closure of the mixing head.
Next time the unit is used just remove the old mixing tip by an anti-clockwise turn of 90° and replace with
a new one.
Attention: Never feed against a mixing tip with cured material!
9.2.4 Change of bottles/canisters
Remove the quick-coupling/bottle adapter from the unit by pulling up the latching. Screw quick-coupling/
bottle adapter on the new bottles/cubitainers and connect as described above. Observe the correct alloca-
tion of the components according to the colour. Always check the filling levels of the containers! They must
not be emptied completely, as otherwise air gets into the system and the unit has to be bled under great
effort. A minimal filling of 2-3 cm has to remain in the containers.

10. Service and Maintenance

Disconnect the unit from mains supply before any service or maintenance!
Cleaning is done with a mild cleaning liquid. Wipe away silicone residues from the unit. Keep coupling plug
and mixing head clean. The cleaning liquid must not get into the unit.
8

11. Troubleshooting

Problem
Cause
Unit does not feed
Mixing head is clogged.
Unit does not produce Cured material in old mixing tip.
Unit does not produce The components A and B have been
interchanged; the silicone has cured
in the system of the unit
Unit does not produce The filling of the silicone containers
has fallen below the minimum filling;
the system has sucked air.
Unit does not feed;
Miniature fuse is defect.
lamp of the operating
button does not light up.
Unit does not produce
The silicone components do not
in a mixing ratio of 1:1.
have the same viscosity.
ENU
Aid
Clean the mixing head carefully by
means of the enclosed cleaning spirals.
Use a different spiral for each drilling,
as otherwise the two components would
mix and cure.
Assemble new mixing tip.
The unit hast to be dismounted
completely and cleaned by an expert.
Unscrew mixing head flange (no. 27)
with a 4 mm Allen key. Remove mixing
heads (no. 3) to the front. Loosen the
union nuts (no. 8a) with a 10 mm spanner
from the hoses and pull off hoses.
Refill silicone material and wait for
approx. 20 minutes until the silicone
has flown into the pipe. Put reception
bowl under the opening of the hoses,
switch unit on and press button until
both components are fed out of the
hoses free of air bubbles.
Reassemble unit in reversed order.
Check micro fuse and exchange
if necessary.
Use material recommended by the
manufacturer (a deviation of about
5 % can be tolerated).
9
Tabla de contenido
loading

Tabla de contenido