5.1

Transport position

Only transport the light in this position.
6.

CLEANING AND DISINFECTION

DANGER
Death hazard from electric shock
Before disinfection cleaning, switch off the power
connection and secure against unintentional activation.
CAUTION
Material damage due to incorrect cleaning
For cleaning, use only products that do not impair
operation of the luminaire.
For cleaning, do not use solvent-based, chlorine-
based or abrasive detergents, because they can crack
plastic parts and cause other damage.
The cleaning agents must be approved for use on
plastics such as PC, PMMA, PA and ABS.
Concentrated disinfectant can damage the luminaire.
For concentration and application times, check the
information provided with the product used.
The wrong cloth may cause scratches.
RECOMMENDED DISINFECTANTS
Bacillol 30 Foam
Dismozon Plus
Kohrsolin Extra
Lysoformin
Mediclean
Meliseptol Foam Pure
Microbac Tissues
Mikrozid Sensitive Liquid
CAUTION
Dirt reduces the light strength
Keep cover clear through regular cleaning.
Only wipe cleaning allowed.
Terralin PAA
Terralin Protect
Virex Tb
CaviCide 1
Gemicidal Bleach
Hexaquart XL
Clean the PA screen with a non-abrasive cleaning cloth
and a suitable cleaning agent.
CAUTION
To minimize the risk of disease transmission, applicable
health and safety regulations and the requirements of the
national hygiene and disinfection authorities must be
observed in addition to these instructions.
7.

SAFETY INSPECTIONS

DANGER
Death hazard from electric shock.
Unplug from the grid.
The power supply cable must be checked at least
once a year for damage.
CAUTION
Maintenance and repairs can be performed only by
qualified electricians.
The corresponding user profile is in Chapter 2 Safety
instructions.

EVERY YEAR:

Check the power supply cable for damage and
replace if necessary.
Check for deformations and cracks in the plastic parts.
Check the load bearing system for distortion or
damage.
Check for loose parts.
8.

DISMANTLING

DANGER
Death hazard from electric shock.
Before dismantling the luminaire, disconnect all pins from
the power supply.