Table 2 Guidance and MANUFACTURER's declaration – electromagnetic IMMUNITY
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment
IMMUNITY test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply i
nput lines IEC 61000-4-11
Power frequency
(50/60Hz) magnetic field IEC 61000-
4-8
for all ME EQUIPMENT and ME SYSTEMS
IEC 60601 test level
Compliance level
±6 kV contact
±8 kV air
±2 kV for
power supply lines
±2 kV for power
±1 kV for
input/output lines
±1 kV line(s)
to line(s)
±1 kV line(s) to
±2 kV line(s)
to earth
<5% UT
(>95% dip in UT)
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT )
(60% dip in UT )
for 5 cycles
70% UT
(30% dip in UT )
(30% dip in UT )
for 25 cycles
<5% UT
(>95% dip in UT )
(>95% dip in UT )
for 5 s
3A/m
NOTE UT is the a.c. mains voltage prior to application of the test level.
61
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
±6 kV contact
If floors are covered with synthetic material, the
±8 kV air
relative humidity should be at least 30%.
Mains power quality should be that of a
supply lines
typical commercial or hospital environment.
Mains power quality should be that of a
line(s)
typical commercial or hospital environment.
<5% UT
Mains power quality should be that of a typical
for 0.5 cycle
commercial or hospital environment.
40% UT
If the user of the device requires
continued operation during power mains inter-
for 5 cycles
70% UT
it is recommended that the device be powered
for 25 cycles
from an uninterruptible power supply or a battery.
<5% UT
for 5 s
Power frequency magnetic fields should be at
3A/m
levels characteristic of a typical location in a
typical commercial or hospital environment.
ruptions,