Error Code Table; Rs 232 Protocol 2 Communication Descriptions; Transmission Channel Characteristics - Agilent Technologies 969-9518 Manual De Instrucciones

Turbo-v 70 pcb 24 vdc
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43

Error Code Table

Tab. 3
CODE
00E000
00E001
00E002
00E003
00E004
00E005
00E006
00E009
00E010

RS 232 Protocol 2 Communication Descriptions

Transmission Channel Characteristics

Tab. 4
levels:
baud rate:
character length:
parity:
stop bit:
protocol:
In this case the value to be assigned to the ADDRESS field must be 80
hex .
Turbo-V 70 PCB 24 Vdc Controller User Manual / 87-900-933-01 (H)
Data Exchange Protocols
DESCRIPTION
No error
Output current > 15 A
Not connected pump
Output current > 1.5 A for 15 sec.
Bearing temperature > 60 °C
Heat exchanger temperature > 60 °C
Normal not reached at run up time
Vp voltage 8V lower than no~ for 5 sec.
Soft start ramp not ended within the expected time
RS 232
9600/4800 programmable by a jumper on the board
8 bits
none
1 bit
master (PC) / slave (converter)
Technical Information
191/208