NOTE:
The compressor can
only be used in con-
junction with the
nebuliser
recommended by
INQUA.
Performing the
inhalation
40
Unwind the power cord (4) on the bottom of the compressor.
Place the compressor (1) on a flat, stable surface.
Plug the power plug into the socket.
WARNING: Danger to health if connection tubings are mixed up
Check carefully to ensure that the other end of the connection tubing
connected to the compressor is attached to the nebuliser.
If several tubing systems are used for different applications at the
same time (e.g., infusion and inhalation), there is a danger that the
tubing systems may be confused. In extreme cases, this can even
result in death.
Place one end of the connection tubing (3) firmly over the air out-
let on the compressor.
Place the other end over the corresponding connector on the
nebuliser (2, see instructions for use of the nebuliser).
Assemble the nebuliser as described in the instructions for use of
the nebuliser.
Attach the mouthpiece or the mask.
Place the nebuliser in the nebuliser holder on the compressor to
fill it.
®
The INQUA
Inhalator is ready for use.
WARNING: Danger to health from inhaling unsuitable medica-
tions or incorrect dosages or incorrect use
Follow the instructions in the package insert for the medication.
Before inhalation, read and comply with the instructions for use
of the nebuliser.
Children, and anyone who requires assistance must be super-
vised constantly by an adult during inhalation therapy. Otherwise,
safe and effective treatment cannot be guaranteed.
CAUTION: Danger of burning if the air outlet is touched directly
during inhalation
The ambient air that passes from the compressor to the connection
tubing heats up the air outlet. As a result, the air outlet can become
very hot. Therefore, avoid direct contact with the air outlet during
inhalation.