Display Indicators

ITEM
DESCRIPTION
Display Update Rate
Type
Pixels
SAFET Y STANDARDS COMPLIANCE
ANSI/AAMI ES 60601-1:2005
CAN/CSA C22.2 No. 60601-1
IEC 60601-1:2005
IEC 62366
IEC 60601-1-6
IEC 60601-1-11
EN/ISO 80601-2-61:2011
Guidance and Manufacturer's Declarations - Electromagnetic Emissions
The ME Equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME Equipment should assure that it is used in such an environment.
EMISSION TEST
COMPLIANCE
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage fluctuations/ Flicker
emissions
IEC 61000-3-3
Guidance and Manufacturer's Declarations - Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME Equipment should assure that it is used in such an environment.
IEC 60601
IMMUNIT Y TEST
TEST LEVEL
Electrostatic
+6 kV contact
discharge (ESD)
+8 kV air
IEC 61000-4-2
+/- 2 kV for power lines
Electrical fast
transient/ burst
+/- 1 kV for input/
IEC 61000-4-4
output lines
+/-1 kV line(s) to
line(s)
Surge
IEC 61000-4-5
+/- 2 kV line(s) to earth
100% dip in mains
Voltage dips, short
voltage for 0.5 cycle
interruptions, and
voltage variations
60% dip in mains
on power supply
voltage for 5 cycle
input lines
30% dip in mains
IEC 61000-4-11
voltage for 25 cycle
1 second
TFT LCD
320 dots x 240 dots
ME Equipment uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any
Group 1
interference in nearby electronic equipment.
Class B
Suitable for use in all establishments, including domestic environments and those directly connected to the public low-voltage power supply
Class A
network that supplies buildings used for domestic purposes.
Complies
COMPLIANCE LEVEL
+6 kV contact
+8 kV air
+/- 2 kV for power lines
+/- 1 kV for input/
output lines
+/-1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
100% dip in mains
voltage for 0.5 cycle
60% dip in mains
voltage for 5 cycle
30% dip in mains
voltage for 25 cycle
Compliance
EMC COMPLIANCE
IEC 60601-1-2:2007, Class B
EN/ISO 80601-2-61:2011, Clause 202.6.2.3, 20 V/m
EQUIPMENT CLASSIFICATION PER IEC 60601-1
Type of Protection
Degree of Protection of Electrical
Shock
Protection against harm from
liquid ingress
Mode of Operation
ELEC T ROM AGN E T I C E N V I RON M E N T - G U I DA N C E
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
12
Class II (AC power)
Internally powered (Battery power)
Defibrillation proof BF-Applied Part
IP22, Protection from ingress of particulates >
than 12.5 mm and protection against vertically
falling water drops when enclosure is tilted at
15 degrees.
Continuous operation
ELEC TROMAGNETIC
ENVIRONMENT - GUIDANCE
39296/9527H-0320