Uline OHAUS DEFENDER 5000 Manual Del Usuario página 4

Báscula digital
Ocultar thumbs Ver también para OHAUS DEFENDER 5000:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MENU NAVIGATION
To enter the Main Menu,
press the
button
User
Menu
from any application
home screen. (See Figure 2)
CHANGING SETTINGS
To change a menu setting, follow these steps:
1.
Enter the Menu – From any application screen, press
the
button. The Main Menu List appears.
User
Menu
2. Select the Submenu – Scroll to desired submenu in
the Main Menu List using the softkey corresponding
to the ▼ icon. Press the softkey corresponding to the
icon to display the Submenu items.
3. Select the Submenu Item – Scroll to the desired
submenu Item using the softkey corresponding to
the ▼ icon. Press the softkey corresponding to the
icon to view the submenu item's settings.
4. Select the Setting – Scroll to desired Setting using
the softkey corresponding to the ▼ icon. Press the
softkey corresponding to the
setting. Press the softkey corresponding to the
icon to return to the previous screen. Press the
softkey corresponding to the
and return to the last active application mode.
CALIBRATION
Three calibration methods are available:
Zero Calibration, Span Calibration and Linearity Calibration.
1.
Make sure that appropriate calibration masses
are available before calibration. See uline.com for
calibration weights.
2. Make sure that the scale base is level and stable
during the entire calibration process.
3. Calibration is unavailable with LFT set to ON.
4. Allow the scale to warm up for approximately five
minutes after stabilizing to room temperature.
5. To abort calibration, press the softkey corresponding
to the "X" icon anytime during the calibration process.
6. When any selection within the GMP menu is enabled,
calibration results are automatically printed.
ZERO CALIBRATION
Zero calibration uses one calibration point. The zero
calibration point is established with no weight on scale.
Use this calibration method to adjust for a different dead
load without affecting span or linearity calibration.
PAGE 4 OF 101
OPERATION
Figure 2
icon to select the
icon to exit menu
User
1.
Long press the
Menu
button to enter main
menu. Press the softkey
corresponding to the
icon to enter the
Calibration submenu. (See Figure 3)
2. When Zero Calibration
is highlighted, press the
softkey corresponding
to the
icon to
initiate. (See Figure 4)
3. Clear the pan and
then press the softkey
corresponding to the
icon. (See Figure 5)
4. The message
"Completed" will
display on screen.
(See Figure 6)
5. Exit Zero Calibration
by pressing the softkey
corresponding to the
icon. (See Figure 7)
6. To return to Main
Menu, press the softkey
corresponding to the
SPAN CALIBRATION
Span Calibration uses one point. This point is established
when a calibration mass is placed on the scale.
NOTE: Span Calibration should be performed
after Zero Calibration.
1.
Long press the
User
Menu
button to enter Main
Menu. (See Figure 8)
2. Press the softkey
corresponding to the
icon to enter the
Calibration submenu.
(See Figure 9)
3. Scroll to Span Calibration using the softkey
corresponding to the ▼ icon.
Figure 3
Figure 4
Figure 5
Figure 6
Figure 7
icon.
Figure 8
Figure 9
0621 IH-8105