The FreeStyle® Lite Test Strip
Apply blood or control solution to only
one test strip sample area per test.
Important: Use only one sample area of the test strip per
test. Do not apply blood to both sample areas. Test strips
may be used only once. Discard used test strips. You can
continue to fill the test strip for up to 60 seconds. Be sure
to reapply the sample to the same sample area.
WARNING:
The cap or vial contains drying agents that may be harmful if
inhaled or swallowed and may cause skin or eye irritation.
Important: The FreeStyle Freedom® Lite meter should only be used
with FreeStyle Lite test strips. Using other test strips with this meter can
produce inaccurate results.
Important Test Strip Information
•
The FreeStyle Freedom Lite system has an operating range of 40° to
104° F (4° to 40° C).
•
Store the test strip package in a cool, dry place between 40° and 86° F
(4° to 30° C).
•
Use test strips only within the system operating temperature range.
•
Keep away from direct sunlight or heat.
•
Store your test strips in their original vial only; never transfer them
to another vial or any other container.
•
Never store individual test strips outside the vial.
•
After removing a FreeStyle Lite test strip from the vial, immediately
close the vial cap tightly.
•
With clean, dry hands, you may touch the test strip anywhere when
removing it from the vial or inserting it into the meter.
9
Call Customer Care Toll Free at 1-888-522-5226 www.AbbottDiabetesCare.com