portable and mobile RF communications equipment and the [MDS3003]/[MDS3003P]
The [MDS3003]/[MDS3003P] is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the [MDS3003]/[MDS3003P] can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the [MDS3003]/[MDS3003P] as recommended below,
according to the maximum output power of the communications equipment.
Rated
maximum
output power of
transmitter W
0.01
0.1
1
10
100
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
 
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
 
0.12
0.38
1.2
3.8
12
-31-
800 MHz to 2,5 GHz
 
0.23
0.73
2.3
7.3
23