Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF
communications equipment and the 04-610-001.
The 04-610-001 is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the
user of the 04-610-001 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the 04-875-001 as
recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of
Rated
transmitter.
maximum
output power
150 kHz to 80
80 MHz to 800
of transmitter
MHz
W
d=1.2
P
0,01
0.12
0,1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be determined
using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people and animals.
24
800 MHz to 2.5
MHz
GHz
d=1.2
d=1.2
P
P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23