¡WARNING!
The steam outlet (13) expels high temperature steam and
could produce serious burns if you don't take precautions.
-
Liquids sterilization: do not open the valve when it cools down till the pressure
gauge (1) indicates 0 bar. Then, set the valve (12) to the «Steam» position.
In both cases, please open the autoclave lid when the pressure gauge indicates 0
bar, as explained in chapter 8.2.
When the autoclave is cooling with the valve closed, some vacuum can be produced
inside the chamber. That can avoid the lid from opening.
In case of not being able to open the lid, break the vacuum by turning the valve (12)
onto «steam» position.
Do not forget to turn the valve back on the «closed» position before the next ste-
rilization.
9 Alarm indicators
Overtemperature indicator (2).
4
5
The safety thermostat has disconnected the heater:
Possible cause 1:
-
The equipment has been switched on without or with not enough water.
Possible solution:
1. Add water until the heater element cover rack, as indicated in chapter 8.1.
2. Press the safety thermostat restart button (19) and begin the sterilization process
once again.
9
10
Possible cause 2:
-
The thermostatic valve did not get closed after the purge phase, and the water
in the sterilization chamber has been evaporated.
Possible solution:
1. It must be solved by a technical service autorized by J.P. Selecta. s.a.u.
Open lid indicator (3)
Possible cause:
-
The indicator is on while the lid is open and must be off when closing it. If it does
not switch off or on when operating, depressurize carefully the autoclave and
contact an authorized technical service of JP Selecta, s.a.u.