7 Installation

7.1 Location

Locate the autoclave near an appropriate power source for the machine.
According to the current legislation, the autoclave should be located in a
way that the safety valve discharge outlet DO NOT point to any people or
could not reach them in case of steam outlet.
Install the autoclave over a flat, stable horizontal surface, suitable for the machine
weight (see chapter 3 «Technical Specifications»), leaving a free space of at least 20
cm around the equipment.
Fix a pressure hose to the outlet (13).
WARNING
If you use the machine for quick depressurization, please fix a pressure hose.
Fix the hose in such a way that cannot move by the effect of the pressure at the
end of the cycle.
Do not hold the hose with the hands. It could reach temperatures higher than 60ºC.

8 Operation

8.1 Preliminary operations for switching on

-
Check that the lateral valve (12) is in the «Closed» position.
-
Place the heater element cover tray at the bottom of the vessel.
-
Fill the sterilization chamber with water until the heater element cover rack.
-
Plug the autoclave to the mains.
-
Press the main switch on (16).
-
Place the material to be sterilized.
-
Close the lid.

8.2 Opening and closing the lid

CLOSING
-
To close, just press over the lid with one hand and turn the knob (15) with the
other hand till the «CLOSED» position.
15
OPENING
-
It is not possible to open the lid if the pressure is different to 0 bar.
-
To open, press the button (6) and turn the knob (15) with the other hand till
the «OPEN» position.
15
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España