8.3 Starting up
Once the operations in chapter 8.1 are made, please do as follows:
13
-
Make sure that the lateral valve (12) is in the «Closed» position.
-
Select the sterilization temperature (121ºC/134ºC) by means of the selector (4).
-
Select the sterilization time with the timer knob (5).
Recommended sterilization temperatures and times:
Material
Liquids
Metallic tools and glass
IMPORTANT
Is advised to use decalcified water.
Distilled water and deionized water produce a chemical reac-
tion which shorten life of the heating elements.
¡WARNING!
The timer has a double scale:
- Blue for equipments connected to a 60 Hz frequency.
- Red for equipments connected to a 50 Hz frequency.
-
The autoclave will begin the purge process by means of which the air is expelled
through the exit (13). At the end of the purge process, the autoclave closes
automatically the steam exit.
-
Once the valve is closed (12), the temperature and pressure increase till they
5
arrive to 121ºC/1 bar or 134ºC/2 bar, according to what it has been selected.
-
When the autoclave reaches the operating temperature, the green indicator lights
on (9) and time begins to deduct from the sterilization time.
-
When the sterilization time finishes, the indicator lights (10) and the autoclave
finishes the sterilization cycle.
8.4 Sterilization end
Once the sterilization time ends, one of these two situations can be carried out:
9
10
-
Solids sterilization: a quick devaporization can be made by setting the valve (12)
to the «steam» position.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Febrero 2020
Temp.
121ºC
134ºC
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Pag.: 22
Time
30'
10'