HealthSmart 04-630-001 Manual De Instrucciones página 24

Ocultar thumbs Ver también para 04-630-001:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF
communications equipment and the 04-630-001, 04-630-006, 04-631-001
and 04-631-006.
The 04-630-001, 04-630-006, 04-631-001 and 04-631-006 are intended for
use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the 04-630-001, 04-630-006, 04-631-001
or 04-631-006 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the 04-630-001, 04-630-006, 04-631-001 or 04-631-006
as recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of transmitter.
Rated maximum
output power of
transmitter W
0,01
0,1
1
10
100
Table 4
150 kHz to 80
80 MHz to 800
MHz
d=1.2 P
0.12
0.38
1.2
3.8
12
24
800 MHz to 2.5
MHz
d=1.2 P
d=1.2 P
0.12
0.38
1.2
3.8
12
GHz
0.23
0.73
2.3
7.3
23

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

04-631-00104-630-00604-631-006

Tabla de contenido