RS Pro True RMS Manual De Instrucciones página 9

Multímetro digital profesional
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Professional True RMS Digital Multimeter/ English
5. Always use a UL recognized fuse of the proper size and value: 800mA/1000V (6.3 x 32mm)
fast blow for the µA/mA ranges and 10A/1000V (10 x 38mm) fast blow for the 10A range.
6. Install the back cover and tighten the screw.
WARNING: To avoid electric shock, do not operate meter until the back cover is securely
fastened to the meter.
Specifications
Accuracy calculated as [%reading + (num. digits*resolution)] at 18°C to 28°C <75%HR
DC Voltage
Range
Resolution Accuracy
600.0mV
0.1mV
6.000V
0.001V
±
(0.5%reading + 5digits)
60.00V
0.01V
600.0V
0.1V
±
(0.8%reading + 5digits)
1000V
1V
AC TRMS Voltage
Range
Resolution
(50Hz to 60Hz)
600.0mV
0.1mV
6.000V
0.001V
±(1.0%reading +
60.00V
0.01V
5digits)
600.0V
0.1V
1000V
1V
(*) Accuracy specified from 5% to 100% of the measuring range, Input impedance: >9MΩ ;
Distorted ,pulsed , triangle or trapezia waveform Accuracy: ±(10%rdg + 10dgt)
Low Pass Filter : Rang 1000V 50/60HZ, ±(1%+20)
60-400Hz ±(3%+20) >3KHz (-3dB)
15
Input
Protection against
impedance
overcharge
>10MΩ
1000VDC/ACrms
Accuracy (*)
Protection against
overcharge
(61Hz to 1kHz)
±(3.0%reading +
1000VDC/ACrms
5dgt)
Professional True RMS Digital Multimeter/ English
AC/DC Voltage(Low Z)
Range
Resolution
Accuracy ±(% of reading + digits)
6.000V
1mV
60.00V
10mV
±(3.0% + 40 digits)
600.0V
0.1V
Input Protection: 600V AC RMS or 600V DC
Input Impedance: Approx.~3kΩ
Accuracy AC+DC function ±(3.5%rdg + 40dgt)
DC Current
Range
Resolution
Accuracy
600.0µA
0.1µA
6000µA
1µA
±(1.0%reading + 5digits)
60.00mA
0.01mA
600.0mA
0.1mA
±(1.0%reading + 8digits)
10.00A
0.01A
±(1.5%reading + 8digits) Quick fuse 10A/1000V
AC TRMS Current
Range
Resolution Accuracy (*) (50Hz to1kHz) Protection against overcharge
600.0µA 0.1µA
6000µA
1µA
±(1.2%reading + 5digits)
60.00mA 0.01mA
600.0mA 0.1mA
10.00A
0.01A
±(1.5%reading + 8digits)
(*) Accuracy specified from 5% to 100% of the measuring range
Distorted ,pulsed , triangle or trapezia waveform Accuracy: ±(10%rdg + 10dgt)
4-20mA% reading
Range
Resolution
Accuracy
-25% to125%
0.1%
±50dgt
Protection against
overcharge
600VDC/ACrms
Protection against overcharge
Quick fuse 800mA/1000V
Quick fuse 800mA/1000V
Quick fuse 10A/1000V
Correspondence
0mA=-25%, 4mA=0%, 20mA=100%,
24mA=125%
16
Tabla de contenido
loading

Este manual también es adecuado para:

Rs-9519bt

Tabla de contenido