CALIBRATION
This scale has been factory calibrated, and does not require calibration prior to use. If
required, the scale can be calibrated. Calibration of your scale is performed using grams. 5000
grams are required for calibration.
Note: Only weights certified and traceable to national standards should be used for
calibration procedures.
Place the scale on a flat surface without movement or
vibration. Power on the scale.
While the lines appear on the display,
press the following key sequence:
g/oz,
, ZERO, ZERO
Note: Sequence must be entered while the lines appear
on the display.
"CODE" will appear on the display.
Once "CODE" appears on the display, enter the calibration
sequence:
g/oz, ZERO, ZERO, g/oz, ZERO, ZERO
"100100" will appear on the display. Press
"RZERO" will appear on the display to prompt the scale to
take the zero reference. Ensure that the weighing pan or
tray is on the scale with no load, then press
The display will count down from 5 to 0.
If the scale detects any movement, it will display the
"UNSTAB" message and the countdown will be reset until
reference to zero is stable. After the countdown is finished
the "LOAD" message will be displayed.
When "LOAD" appears on the display, place the maximum
capacity of 5000 grams of certified weight on the scale.
Once the weight is on the scale, press
The display will count down from 5 to 0.
If the scale detects any movement, it will display the
"UNSTAB" message and the countdown will be reset until
the weight is stable. At the end, a calibration factor will be
displayed for 5 seconds then display "OFF". A calibration
factor will appear.
Remove weight from the scale and power it off.
P/N UM3400KL_3401KL_3400KG_3400KG-C REV20200203_web
ACTION
to accept.
.
11
DISPLAY SHOWS
'-----'
"CODE"
"100100"
"RZERO"
Countdown from 5 to 0,
followed by "LOAD".
"LOAD"
Countdown from 5 to 0,
followed by a calibration factor.