49
Note: because SpO
probe measurements are statistically distributed, only
2
about two-thirds of SpO
2
±Arms of the value measured by a CO-OXIMETER.
B. B. Pulse rate measurement
Range:30 bpm~250 bpm
Error: ±2 bpm or ±2% (select the larger)
Resolution: 1bpm
C. Optical sensor: red light(wavelength: 660 nm, output power less than 6.65
mW) infrared light(wavelength: 880 nm, output power less than 6.75 mW).
Optical sensors are light-emitting components that affect other medical
devices that use this wavelength range. This information may be useful to
clinicians who perform optical therapy.
D. Error in weak filling condition: SpO
when pulse-filling ratio is 0.4%. SpO
is 30 bpm~100 bpm, pulse rate error is ±2 bpm; when measuring range is
100 bpm~250 bpm, pulse rate error is ±2%.
Note :
• The optional probe of the Sphygmomanometer is an integrated SpO
probe;
• The service life of the integrated SpO
Cuff:
There are several suitable cuffs(range of limb circumference,middle of upper limb)
the range of limb circumference is 6-11 cm
the range of limb circumference is 10-19 cm
the range of limb circumference is 18-26 cm
the range of limb circumference is 32-43 cm
probe measurements can be expected to fall within
and pulse rate can be shown correctly
2
error is ±4%; when measuring range
2
probe is three years.
2
ENGLISH
probe, the measuring part is integrated with the
2