Alarms And Self-Diagnosis - Phcbi MDF-DU302VX Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

ALARMS AND SELF-DIAGNOSIS

Warning
The cooling performance is significantly reduced. The chamber temperature may get higher
:
considerably. Take some precautions for the storage items immediately (transferring the storage items
to another freezer or placing of dry ice wrapped in newspaper in the chamber) except when the cause is
clear and the chamber temperature can be recovered soon.
Contact our sales representative or agent after turning off the power switch.
LCD touch panel
Message display field
Warning: Temp Control Failure.
W01: Power Failure.
Warning: Temp Control Failure.
W02: Compressor 'A' Temp Abnormal.
Warning: Temp Control Failure.
W03: Compressor 'B' Temp Abnormal.
Warning: Temp Too High.
W04
Warning: Temp Too Low.
W05
Warning: Temp Control Failure.
W08: Temperature Controller Failure.
Warning: Temp Control Failure.
W09: Temperature Sensor Error.
Warning: Temp Control Failure. *2
W10: Temperature Sensor Error.
Warning: Temp Control Risk.
W11: Compressor 'A' Control Failure.
Warning: Temp Control Risk.
W12: Compressor 'B' Control Failure.
Warning: Temp Control Risk.
W13: Compressor 'A' Control Failure.
Warning: Temp Control Risk.
W14: Compressor 'B' Control Failure.
1: The compressor stops in the case of W02 or W03.
*
2: The compressor runs continuously in the case of W09 or W10.
*
The compressor stop has a priority over the continuous running if the above two errors come up at one time.
3: The chamber temperature is not displayed in the case of W08
*
Alarm
Cooling performance may decline and the temperature of the chamber may rise. Wait for the
:
recovery of chamber temperature if the temperature change is temporary resulting from user operation.
For other cases, failure or chamber temperature rise may cause if this status continues.
Take some precautions for the storage items (transferring the storage items to another freezer or
placing of dry ice wrapped in newspaper in the chamber).
Contact our sales representative or agent.
LCD touch panel
Message display field
Alarm: Temp Too High.
A04
Alarm: Temp Too Low.
A05
Alarm: Temp Control Risk.
A06: Cooling Circuit 'A' Abnormal.
Alarm: Temp Control Risk.
A07: Cooling Circuit 'B' Abnormal.
Alarm: Temp Control Risk.
A08: Fan Motor 'A' Abnormal.
Alarm: Temp Control Risk.
A09: Fan Motor 'B' Abnormal.
Situation
The battery switch for power failure alarm is
ON, and under any of the following
conditions.
During a power failure
Power switch is OFF
Power supply cord is disconnected.
*1
Compressor 'A' Temp Abnormality.
*1
Compressor 'B' Temp Abnormality.
If the chamber temperature exceeds the
set temp. + the set value of High Alarm.
If the chamber temperature falls below the
set temp. - the set value of Low Alarm.
When communication between LCD touch
* 3
panel and control substrate is died out or
unstable.
*2
If the thermal sensor is disconnected.
If the thermal sensor is short-circuited.
Compressor 'A' does not start-up. Or,
diagnostic sensor failure.
Compressor 'B' does not start-up. Or,
diagnostic sensor failure.
Compressor
'A'
does
diagnostic sensor failure.
Compressor
'B'
does
diagnostic sensor failure.
Situation
If the chamber temperature exceeds the
set temp. + the set value of High Alarm.
If the chamber temperature falls below the
set temp. - the set value of Low Alarm.
The performance of the cooling circuit "A"
gets worse.
The performance of the cooling circuit "B"
gets worse.
Fan motor "A" failure.
Fan motor "B" failure.
Buzzer
Intermittent
tone
Intermittent
tone
not
stop.
Or,
not
stop.
Or,
Buzzer
Intermittent
tone
Remote
Alarm & safety
alarm
Power failure
alarm
Alarm
Compressor 'A' Temp
mode
Abnormality
*1
Compressor 'B' Temp
Abnormality
*1
High Alarm
Low Alarm
Communication error
Temperature Sensor
Alarm
disconnected
*2
mode
Temperature Sensor
short-circuited
*2
A Compressor
Start-up failure.
B Compressor
Start-up failure.
A Compressor Stop
failure.
B Compressor Stop
failure.
Remote
Alarm & safety
alarm
High Alarm
Low Alarm
Cooling circuit
abnormality
Fan Motor 'A'
Abnormality
Fan Motor 'B'
Abnormality
47
Tabla de contenido

Solución de problemas

loading