BASIC PARAMETERS
4.
Enter the required settings. Press Apply to save
the settings. The display returns to the Tools
screen.
Settings
•Control:
Choose the operation control mode between the NORMAL (normal control) or the ECO (eco control). By
holding the Control slide key and sliding it right, operation control mode is changed to ECO. Factory
setting: NORMAL.
•Compressor Delay:
This controls the time from turning the unit ON until starting its compressor. This unit requires a lot of
electrical power at the moment its compressor starts. When several units are kept in the same room,
adjust this setting so that the compressor delay times of the different units are staggered as this will
prevent the simultaneous restart of all compressors after a power failure. Settable range: 3 minutes~15
minutes, factory setting: 3 minutes.
Note: This unit has two compressors A and B. The compressor whose cumulative operating time is
shorter than the other one, starts first, after 1 minute the other starts.
5.
Press Top key to return to the Top screen.

Setting key lock

1.
Press Menu key to access the Menu screen.
2.
Press Key Lock key to access the Key Lock
screen.
26