BASIC PARAMETERS
Setting Temperature and temperature alarms
Set the Temperature, High Temperature Alarm and Low Temperature Alarm for normal operation using
the following procedure. The unit automatically starts operation using these settings after the power is
switched on
1.
Press Menu key to access the Menu screen.
2.
Press Set key to access the Temp. Setting
screen.
3.
Enter the required settings. Press Apply to save
the settings. The display returns to the Menu
screen.
●Parameter settings
• Temperature: Set value of chamber temperature.
Settable range: -90
• High Alarm: When the chamber temperature exceeds the High Alarm set temperature (= the set
temperature + the set value of High Alarm)*, the High Alarm is activated.
Settable range: +5
• Low Alarm: When the chamber temperature falls below the Low Alarm set temperature (= the set
temperature - the set value of Low Alarm)*, the Low Alarm is activated.
Settable range: -40
24
o
C~-50
o
C, Control range: -86
o
C~+40
o
C, factory setting: +10
o
C~-5
o
C, factory setting: -10
o
C~-50
o
C, factory setting: -80
o
C.
o
C.
o
C.