Important Control Solution Information - Fora D30 Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Important Control Solution Information

• Use only FORA control solutions with your monitor.
• Do not use the control solution beyond the expiration date or 3
months after first opening. Write the opening date on the control
solution vial and discard the remaining solution after 3 months.
• It is recommended that the control solution test should be done at
room temperature (68 ° F to 77 ° F / 20 ° C to 25 ° C). Make sure your
control solution, monitor, and test strips are at this specified
temperature range before testing.
• Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
• Store the control solution tightly closed at temperatures between
36 ° F and 86 ° F (2 ° C and 30 ° C). Do NOT freeze.
55

Capítulos

Tabla de contenido
loading

Tabla de contenido