Disc not spinning
Test result value on the
display is red
Low Nitrate or Nitrate-N
results
Trouble connecting to
Bluetooth device
Trouble connecting to
computer by USB
Results not printing
TROUBLESHOOTING WITH THE METER CHECK DISK
Do not attempt to separate the components of the Meter Check Disk (Code 1705). The
Meter Check Disk consists of a disk with a permanently attached cover.
Do not fill the Meter Check Disk with water. Water is not used during the meter check
procedure.
Range Check Procedure
The Meter Check Disk is used to verify the performance of a meter. It is not used to
recalibrate the meter. Meters are calibrated at the time of manufacture. The meter is
performing satisfactorily if readings using the Meter Check Disk are within the ranges
provided on the Meter Check Disk pouch. Range specifications are specific to the disk
identified by the serial number on the pouch. The range specifications will vary from disk to
disk. The exact readings from a specific disk may vary from meter to meter.
1. Follow the Cleaning procedure on page 21 to clean the light chamber and optic lenses.
2. Tap
select Disk Series.
3. Tap MCD to select Meter Check Disk.
4. Tap
to return to the test screen.
5. Remove the Meter Check Disk (Code 1705) from the foil pouch. DO NOT remove the
black cover from the disk.
6. Insert the Meter Check Disk. Close the lid.
Lid open, meter not powered on,
low battery, disk or disk cover
pressed down too tightly on hub
Fast electrical transients may
disrupt operation of the Spin
Touch
meter
™
Results are out of range of
reagent system
Recent treatment with chlorine
neutralizer containing sodium
thiosulfate interferes with test
reaction
Bluetooth not enabled
Too many Bluetooth devices near
the meter
Printer power is ON and Fast
Printer Connection is ON
Broken connection
Printer not on
The Spin Touch
will only print
™
to the Mobile Bluetooth Printer
(Code 5-0066)
19
Close lid, power on meter, charge
the battery or plug meter into a
stable power source, remove the
disk/disk cover and place back in
the meter more gently
Restart the test to resume normal
operation
Dilute sample. Retest for out of
range test factor. (All factors
except pH.)
Retest in 2 – 3 days
Enable Bluetooth
Have only one device near the
meter
Turn printer power OFF. Or turn
Past Printer Connection OFF.
Press and hold power button for
1 second.
Turn printer on
Connect to Mobile Bluetooth
Printer (Code 5-0066)