Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The WiSe™ 26" HDTV Surgical Display and WiSe™ HDTV Transmitter are intended for use in the
electromagnetic environment specified below. The customer or the user of the WiSe™ 26" HDTV Surgical
Display and WiSe™ HDTV Transmitter should ensure that they are used in such an environment.
Immunity Test
Electrostatic
Discharge
(ESD)
IEC61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
2kV common mode
Voltage
<5% Ut (>95% dip
dips, short
in Ut) for 0.5 cycle
interruptions
and voltage
40% Ut (60% dip in
variations on
power supply
input lines
70% Ut (30% dip in
Ut) for 25 cycles
IEC61000-4-11
<5% Ut (>95% dip
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE: Ut is the AC mains voltage prior to application of the test level.
IEC 60601 Test
Level
6kV contact
8kV air
2kV for power
supply lines
1kV for input/
output lines
1kV differential
mode
Ut) for 5 cycles
40% Ut (60% dip
in Ut) for 5 cycles
in Ut) for 5 sec.
3.0 A/m
Compliance
Level
6kV contact
ceramic tile. If floors are covered with
8kV air
synthetic material, the relative humidity
2kV line to
ground
1kV line to line
1kV differential
mode
2kV common
mode
<5% Ut (>95%
dip in Ut) for 0.5
cycle
70% Ut (30%
interruptions, it is recommended that
dip in Ut) for 25
the Wireless Transmitter be powered
cycles
from an uninterruptible power supply
<5% Ut (>95%
dip in Ut) for 5
sec
3.0 A/m
should be at levels characteristic of a
typical location in a typical commercial
Electromagnetic Environment--
Guidance
Floors should be wood, concrete, or
should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the transmitter requires continued
operation during power mains
or a battery.
Power-frequency magnetic fields
or hospital environment.