List of Error Codes

The alarm lamp flashes when a medium-priority alarm occurs, and lights steadily when a low-priority
alarm occurs.
To clear an alarm, press any button.
If a low-priority alarm and a medium-priority alarm occur at the same time, the medium-priority alarm
is displayed.
If the alarms are the same priority level, the error code of the alarm that occurred first is displayed.
However, to prevent battery consumption when powered by battery, a low battery error (E40 error) is
always given priority.
Example: E2
SYSTEM
Error
Priority
code
E9
Medium Internal hardware error
NIBP
Error
Priority
code
The cuff tube is not connected
E1
Low
Air is leaking from the cuff.
Did not inflate properly because the arm
or body moved during measurement.
Moved body or arm during
measurement, or talked.
The cuff is not applied correctly.
The sleeve is rolled up and is
E2
Low
compressing the arm.
Measurement time has exceeded
specified time.
Specified time: 165 seconds
28
Description
Contact your OMRON retail dealer or
distributor.
Description
Firmly connect the cuff tube.
Replace with an OMRON GS CUFF that
does not leak.
Have the patient not move the arm or body,
and repeat measurement.
Have the patient not talk or move, and
repeat measurement.
Correctly apply the cuff.
Remove the garment and re-wrap the cuff.
The measurement time exceeds the
expected time, so the measurement was
ended in order to avoid patient discomfort.
There is a possibility that measurement is
being repeated over and over due to air
leaking from the cuff.
Points to check
Points to check