Technical Data - boso medistar S Manual Del Usuario

Ocultar thumbs Ver también para medistar S:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Technical data

Measurement principle:
oscillometric
Measurement range:
40 to 240 mmHg
40 to 200 pulses/min
Memory store: 30 measurements
Display: LCD
Operating conditions:
ambient temperature: 10 – 40 °C
relative humidity: 10 – 85 %
Transport and storage conditions:
Temperature -5 – 50°C
relative humidity: max. 85 %
Power supply:
DC 3V (2 x 1.5 V microcells alkali manganese,
IEC LR 03)
Battery check:
indicator symbol in display
22
Weight: 0,15 kg
Classification: Type BF
Clinical Test (DIN 58130):
Accuracy complies with EN 1060 part 3
Maximum deviation of cuff pressure: ± 3 mmHg
Maximum deviation of pulse rate: ± 5 %
Dimensions (W x D x H):
75 mm x 80 mm x 65 mm