IRREGULAR HEARTBEAT DETECTOR (IHB)
The appearance of the
icon
indicates that a pulse irregularity
consistent with an irregular heartbeat
was detected during measurement.
Usually this is not a cause for concern.
However, if the symbol appears often,
we recommend you seek medical
advice. Please note that the device does
not replace a cardiac examination, but
serves to detect pulse irregularities at an
early stage.
Movement, shaking or talking during the measurement can result
in pulse irregularities that may cause the appearance of this icon.
Therefore, it is of great importance to not move or talk during
measurement.
To determine the presence of an irregular heartbeat, the average of
the heartbeat intervals is calculated with the first 3 normal effective
heartbeat values. It is important to note that the average is not a
strict mathematical averaging of all recorded intervals. At least 3
beats with 25% or greater difference from the average heartbeat
interval will generate the
Important Information:
This blood pressure monitor is not designed for use by people with
arrhythmias nor for diagnosing or treating an arrhythmia problem. As a
safeguard, we recommend that if you have arrhythmias such as atrial
or ventricular premature beats and atrial fibrillation or any other special
conditions you should check with your physician before using your blood
pressure monitor.
Irregular
Heartbeat
Detector
icon on the screen.
25

USING THE MEMORY FUNCTION

This monitor can be used by 2 users. Each user can
store up to 120 measurements.
This monitor features an advanced memory mode to
provide you with a variety of options for reviewing your
measurement history, including AM/PM Averaging.
Reviewing your morning (AM) and nighttime (PM) blood
pressures can provide important information about how
your blood pressure changes throughout the day.
Displays average of last three readings.
Displays average of last 3 morning
readings.
(AM is defined as 4:00 AM – 12:00 PM).
Displays average of last 3 nighttime
readings.
(PM is defined as 6:00 PM – 2:00 AM).
Indicates reading is an average of a
Triple Check measurement. Press the
button to review individual readings that
make up the Triple Check average.
Note:
A Triple Check average and the 3 measurements that make
this average, count as 4 readings when stored in memory.