VI MOUNTING ON A SOCKET (continuation)
C
Screw the external ring nut (b) on the valve seat (c). Caution : Left-hand thread
D
Place the external gasket (d) on the valve body (a)
E
Screw the valve body (a) fitted with its gasket into the valve seat (c) . Caution : Right-hand thread, a quarter turn
C
VII MANUAL / AUT OMATIC FUNCTION FOR AIR EXPULSION
The 1S163 valve is designed to allow air explusion, either automatically or manually. The selected function is determined by the positi on of the inner
membrane in the valve body. The 1S163 valve is factory set in the "MANUAL EXPULSION" position. To switch from one position to the other, follow
the instructions below:
MANUAL EXPULSION => AUTOMATIC EXPULSION
Unscrew the plug (1) and remove it from the valve
a)
b) Remove the membrane (3)
c) Mount the membrane (3) according to the diagram
d) Place the plug (1) again
e) Press the flap (4) to allow air to flow out.
AUTOMATIC EXPULSION => MANUAL EXPULSION
a) Unscrew the plug (1) and remove it from the valve
body (2)
b) Remove the membrane (3)
c) Mount the membrane (3) according to the diagram
d) Place the plug (1) again
e) Press the flap (4) to allow air to flow out.
VIII RECYCLING
The 1S163 valve includes parts in plastics, silicone, aluminium and steal. Each of these items must be recycled according to the laws i n force.
IX DISCLAIMER
The manufacturer's liability for this product is not involved if the product is not properly used or if the instructions are not followed.
No liability can be accepted for damages that are caused by the interaction of this valve with other components of the prosth esis.
X DECLARATION OF CONFORMITY
This product complies with the requirements of directive 93/42/CEE for medical devices. According to the classification crite ria (Annex IX of the
directive), the product belongs to class I. Moreover, the manufacturer declares to comply with Annex VII of the directive for EC marking procedures.
D
c
b
2
2
d
a
a
1
b
a
1
b
6, rue de la Redoute - 21850 - SAINT APOLLINAIRE -FRANCE
Tél: 33 (0)3 80 78 42 10 - Fax: 33 (0)3 80 78 42 15
c
3
3
c
3
3
c
PROTEOR
[email protected] www.proteor.com
1S16399-1216 4/8
E
a
1
d
e
4
1
d
e
4
Tabla de contenido
loading

Tabla de contenido