HOSPIVAC – MEDIVAC EVISA E25
ALARMS
10.1. For the MILLENIUM
Alarms appear on the display and are:
Disjunction,
Oil level (option),
Network pressure (option).
10.2. For the CYCLIC 2020
When the LED on the
key is lit, an alarm has occurred.
Fault on concerned vacuum pump, may be in connection with:
Constant
Flashing
Upper limit fault on concerned vacuum pump.
Flashing slowly
Regulation pressure fault
Constant
Network pressure fault
Acknowledgement: press
and
10.3. For the PROCOM-2
For more detailed technical information concerning the control mode, consult the PROCOM-2 manual.
CLEANING AND MAINTENANCE
IMPORTANT:
For this operation, use all necessary protective measures (gloves, mask, overalls etc.) then
destroy contaminated consumables (antibacterial cartridges, condensates etc.) in
a o di g ith that ou t 's appli a le la s egulatio i i e ato .
Maintenance is to be entrusted to trained and approved personnel by the exploiting, with
all necessary knowledge and acquainted with best practice.
Not to carry out maintenance operations when the device is in use.
519991-06
Circuit breaking (disjunction)
Oil level (optional).
(parameter 4) = -500 mbar
(parameter 5) = -370 mbar
simultaneously to acknowledge the alarm once read.
E300
If the spare parts used during maintenance operations are not corresponding to the
manufacturers' requirements, the manufacturer declines all responsibility in the event of
an incident.
The modification of the central units is forbidden.
Modification of the device is forbidden without the approval of the manufacturer.
If the central is modified following manufacturer agreement, a control and a suitable test
must be realized to make sure that the central is always safe to use.
To avoid risk of damage or prevent equipment become dangerous is essential use the compatible
components provide by the manufacturer.
FILTRATION
- bleed PYREX vessel
- change main anti-bacteria filter
- change secondary anti-bacteria filter (option)
MAIN and SECONDARY RECEIVERS
- general inspection of condition
- call in an outside inspection body
ELECTRONICS CABINET
- clean and dust using a soft brush
- check the connections in the cabinet(s)
- check the connections at motor terminals
- operate the mechanical back-up systems
- check the intensity of the motors
VACUUM PUMPS
EXHAUST VESSEL
- Empty as necessary
VACUUM HOSE
- Check general condition
- Change
NON RETURN VALVE
- Check tightness
PRESSURE INDICATOR (electronic sensor, vacuum gauge)
- Check measures for correctness
NETWORK ALARM
- Check correct working order
EXHAUST HOSE
- Check the general condition
- Change
Monthly
Quarterly
Annually
2 years
5 years
Regular inspection
#
According to use
#
According to legislation
#
#
#
#
#
Refer to the start-up and maintenance manual
Regular inspection
#
#
#
#
#
#
#
20