MAG1000 is intended for use in the electromagnetic environment specified below. The
customer or the user of MAG1000 should assure that is used in such environment.
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
Electrical
fast
transient/burst
IEC 61000-4-4
Impulses
EN 61000-4-5
Voltage dips, short
interruptions
and
voltage variations on
power supply input
lines
IEC 61000-4-11
Mains
power
electromagnetic field
EN 61000-4-8
IACER S.r.l.
Electromagnetic immunity
Test
level
EN
60601-1-2
6kV contact
8kV air
± 2kV for power
supply lines
1kV
differential
mode
< 5% U
T
(>95% dips of U
)
T
per 0,5 cycles
40% U
T
(60% dips of U
)
T
per 5 cycles
70% U
T
(30% dips of U
)
T
per 25 cycles
< 5% U
T
(>95% dips of U
)
T
per 5 seconds
3 A/m
20 di 23
Compliance level
6kV contact
8kV air
± 2kV per power
supply lines
1kV
differential
mode
< 5% U
T
(>95% dips of U
)
T
per 0,5 cycles
40% U
T
(60% dips of U
)
T
per 5 cycles
70% U
T
(30% dips of U
)
T
per 25 cycles
< 5% U
T
(>95% dips of U
)
T
per 5 seconds
3 A/m
MNPG250 Rev. 01 Ed. 02/05/2018
Electromagnetic
environment
–
guidance
Floors
should
be
wood, concrete or
ceramic tile. If floor
is
covered
with
synthetic
material,
the relative humidity
should be at least
30%
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
If the user of the
MAG1000 requires
continued operation
during power mains
interruptions,
it
is
recommended
that
MAG1000
be
powered
from
an
uninterruptible
power supply or a
battery.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.